The CDT 3200 transformer used for testing according to MIL STD 461 (CS101, CS106, CS109), DO 160 (Section 16 and Section 18), ISO 11452-10, Chrysler CS 11809, Ford FMC 1278, DaimlerChrysler DC-1061, EMC-CS-2010JLR V1.1, MIL-STD-704, Mitsubishi ES-X82115, SAE J1113-2, Tata TST/TS/WI/257.
The CDT 3200 transformer has a frequency range from 30 Hz to 250 kHz. The winding ratio is 2:1. The secondary winding can handle up to 50 A AC or DC without the transformer going into saturation.
The transformer may be used as a 4 Ω primary and 1 Ω secondary or 2.4 Ω primary and 0.6 Ω secondary or 2 Ω primary and 0.5 Ω secondary.
Frequency range: | 30 Hz – 250 kHz |
Turns ratio: | 2:1 |
Primary Impedance: | 4x the secondary impedance |
Audio power: | 200 W |
Secondary current AC+DC nominal: | 50 Ap |
Secondary current AC+DC usable: | 60 Ap |
Primary Inductance (unloaded): | Approximate 5.1 mH |
Secondary Inductance (unloaded): | Approximate 1.3 mH |
Eut Connectors: | Wing terminal |
Primary, AE connectors: | 4 mm safety socket |
Dimensions housing with terminals maximum(W x H x D): | 220 x 225 x 283 mm |
Weight: | 13.0 kg |