Calibration-Pulse Generator acc. CISPR 16 for Band A, B, C, D. To calibrate the pulse response of EMI receivers.
Pulse repetition frequency main generator 0.1 – 200 Hz (aux. generator up to 1 MHz) with IEEE-488 Interface.
Datasheet IGUU 2918
Pulse repetition frequency main generator 0.1 – 200 Hz (aux. generator up to 1 MHz) with IEEE-488 Interface.
Datasheet IGUU 2918
IGUU 2918 Technical Data
Main Generators | |
Pulse frequency range | 0.1 Hz – 200 Hz in 0.1 Hz steps |
Pulse frequency accuracy | 2.3 × 10-3 Hz |
Pulse shape | Almost rectangular |
Pulse duration | Band A: 270 ns; Band B: 6 ns; Band C/D: 0.3 ns; |
Pulse polarity | positive/negative (switchable) |
Pulse level | (CISPR weighting Q.P.): Band A / B 10 – 70 dB, in 1 dB – steps; Band C / D 0 – 60 dB, in 1 dB – steps |
Max. Output Voltage acr. 50 Ω | 80 V – 110 V |
Pulse Area at 60 dB attenuator setting | Corresponds to CISPR 16 Standard equivalent to 1 mVrms across 50 Ω with standardized Q.P. – Detector. |
Frequency flatness | Band A / B: typ. + / – 0 dB to – 0.5 dB at upper frequency limit; Band C: typ. + / – 0.5 dB 30 – 100 MHz (100 MHz is reference frequency for level calibration). Level increase between 200 MHz and 800 MHz + 0.5 to + 1 dB, 800 – 1000 MHz decrease to + / – 0 dB to – 1 dB. |
Level Accuracy | |
Band A / B | + / – 0.5 dB at 50 kHz, 1 MHz |
Band C / D | + / – 0.8 dB at 100 MHz. |
Auxiliary Generators | |
Pulse frequency ranges | Band A: 0.1 Hz – 2 kHz; Band B: 0.1 Hz – 50 kHz; Band C/D: 0.1 Hz – 1000 kHz; 0.1 Hz – 999.9 Hz, 0.1 Hz steps; |
Pulse frequency accuracy | 2.3 × 10-3 Hz |
Pulse duration (values for orientation) | Band A: 260 ns; Band B: 6.5 ns; Band C / D: 500 ps |
Pulse level | (CISPR weighting Q.P.): Band A / B 60 dB fixed level; Band C / D 40 dB fixed level |
Frequency flatness | Band A / B same as 1. Main Generators Band C (D): typ. + / – 0.75 dB 30 – 1000 MHz |
Dimensions (W x H x D): | 447 mm x 180 mm x 460 mm |
Weight | 12 kg |