HHS 5210-100 2.5 – Helmholtz Coils

Schwarzbeck HHS 5210 Helmholtz Coil
Schwarzbeck HHS 5210 Helmholtz Coil
The Schwarzbeck HHS 5210-100 2.5 Square Helmholtz Coils are expertly designed to generate precise magnetic fields over a frequency range from DC to 10 kHz. Each coil features 100 turns and can handle a maximum coil current of 20 A for 5 minutes, achieving a maximum magnetic field strength of 2900 A/m at a 0.4 m coil distance. With an inner side length of 1.0 m and a maximum coil separation of 0.70 m, these coils offer flexibility for various testing and calibration setups. The inductance per coil is 31.3 mH, and the pair has a resonance frequency greater than 25 kHz, ensuring high performance and stability. Optimized for best uniformity at a 0.566 m coil distance, the magnetic field strength per 1 A coil current is 124.66 A/m. These coils are ideal for precise magnetic field calibration and immunity testing, providing exceptional accuracy and reliability.

Number of turns:100
Maximum Coil Current:20 A (5 min.)
Nominal Coil Current:15 A continuous
Inner side length of the coil:1.0 m
Electrical side length of the coil:1.04 m
Maximum Coil Separation:0.70 m
Usable Frequency Range:DC – 10 kHz
Inductance (per Coil):31.3 mH
Resonance Frequency (Pair of Coils):>25 kHz
Weight:56 kg
Terminals:4 mm female
The following data is valid for 0.4 m coil distance:
Coil Separation acc. IEC 1000-4-8:0.40 m
Max. Magn. Field Strength:2900 A/m (5 min.)
Nominal Magn. Field Strength:1746 A/m continuous
Magn. Field Strength, 1 A Coil Current:145.54 A/m (Coil Dist. 0.4 m)
The following data is valid for 0.566 m coil distance:
Coil Separation for best uniformity
along an axes through the coil center:
0.566 m
Magn. Field Strength, 1 A Coil Current:124.66 A/m (Coil Dist. 0.566 m)

Datasheet HHS 5210-100 2.5