The HHS 5206-8 is particularly suited for calibrating magnetic field probes and sensors. Its high temperature-proof Teflon-coated wire ensures durability and reliability, while the linear relationship between coil current and magnetic field strength allows for precise field strength calculations. The coils can accommodate a maximum cubic Device Under Test (DuT) size of 32.5 x 32.5 x 32.5 cm, providing versatility for various testing setups.
This model also supports the NFCN 9732 85 kHz compensation network, which reduces total impedance and allows continuous currents up to 15 Arms at operating frequencies of 80-90 kHz, making it ideal for immunity test inductive charging at 85 kHz. The HHS 5206-8’s robust design and easy installation process make it an excellent choice for high-precision magnetic field generation and calibration. Special care should be taken by individuals with magnetic-field-sensitive implants when using this device.
Number of turns: | 8 |
Maximum Coil Current: | 55 A (5 min.) |
Nominal Coil Current: | 34 A continuous |
Coil Spacings (centered): | 300 mm |
Max. Magn. Field Strength: | 1060 A/m (5 min.) |
Nominal Magn. Field Strength: | 650 A/m continuous |
Magn. Field Strength, 1 A Coil Current: | 19.23 A/m; 145.68 dBµA/m |
Coil Diameter: | 600 mm |
Max. Cubical shaped DuT: | 32.5 x 32.5 x 32.5 cm |
Mechanical Dimensions: | 0.64 m x 0.79 m x 0.42 m |
Terminals: | 4 mm female |
Inductance (Single Coil): | 91 µH |
Inductance (Coil Pair): | 200 µH |
Usable Frequency Range: | DC – 800 kHz |
Resonance Frequency: | > 1500 kHz |
Weight: | 17.0 kg |
Recommended accessories: | NFCN 9732 85 kHz |