The Schwarzbeck HHS 5201-6 Helmholtz Coil is designed to generate precise magnetic fields over a wide frequency range from DC to 5 MHz. With a resonance frequency of over 8 MHz, it ensures high-performance operation in various applications. Featuring 6 turns and a maximum coil current of 40 A for 1 minute, it can achieve a magnetic field strength of up to 2860 A/m for short durations. The coil allows for flexible spacing options between 50 mm and 80 mm. The compact design, with mechanical dimensions of 290 x 220 x 130 mm and a weight of 2.8 kg, makes it easy to integrate into various testing setups. Ideal for high-precision calibration and magnetic immunity testing, the HHS 5201-6 offers reliable performance and exceptional versatility.
Number of turns: | 6 |
Maximum Coil Current: | 40 A (1 min.)15 A (15 min.) |
Nominal Coil Current: | 12 A continuous |
Coil Spacings (centered): | 50 … 80 mm |
The following data refers to a coil separation of 60 mm |
Max. Magn. Field Strength: | 2860 A/m (1 min.) |
Nominal Magn. Field Strength: | 860 A/m continuous |
Magn. Field Strength, 1 A Coil Current: | 71.55 A/m |
Coil Diameter: | 120 mm |
Mechanical Dimensions: | 290 x 220 x 130 mm |
Terminals: | 4 mm female |
Inductance (Single Coil): | 8 µH |
Inductance (Coil Pair): | 18.7 µH / 50 mm; 18.2 µH / 60 mm; 17.9 µH / 70 mm |
Usable Frequency Range: | DC – 5 MHz |
Resonance Frequency: | > 8 MHz |
Weight: | 2.8 kg |
Datasheet HHS 5201-6