CAL EMCL – Calibration Jig for EM Coupling Clamp

CAL EMCL - Calibration Jig for EM Coupling Clamp
CAL EMCL – Calibration Jig for EM Coupling Clamp

Calibration Jig CAL EMCL with opened injection clamp EMCL 6146
Calibration Jig CAL EMCL with opened injection clamp EMCL 6146

100 Ω series resistor with N-female / N-male connectors
100 Ω series resistor with N-female / N-male connectors
The Calibration Jig CAL EMCL can be used to calibrate the coupling factor and the impedance Zin and the decoupling factor of the EM Coupling Clamp EMCL 6146 as well as for test level setting acc. to IEC / EN 61000-4-6 (Immunity to conducted disturbances, induced by radio-frequency fields). Further, the decoupling clamp FTC 101 B can be characterized (Zin, decoupling a).

Usable Frequency Range:10 kHz … 1000 MHz
Frequency Range, normative:150 kHz … 80 MHz
Standards:EN / IEC 61000-4-6
Nominal Impedance Zref‘:247.6 Ω
Impedance range Zref‘:~235 … 258 Ω
Center height of the inner conductor above ground:~62 +/- 12 mm
Diameter inner conductor:4 mm
Connector:N-female
Length x Width x Height:1066 x 500 x 133 mm
Material:Aluminium
Weight:5.8 kg

Datasheet CAL EMCL