DC-Block 500 - DC-blocking capacitor
DC-Block 500 – DC-blocking capacitor
To meet the requirements for measurements with a broadband artificial network (BAN) in the automotive sector, as specified by ISO 11452-7, the isolating capacitor “DC-Block 500” has been developed. It is housed in a robust enclosure and features a BNC socket and a BNC plug, allowing it to be easily integrated into an existing measurement setup.

Precise impedance measurements at high frequencies are always challenging. Therefore, insertion loss is measured instead within a 50 Ω system. A 5 Ω resistor would result in an attenuation of 0.42 dB. If the attenuation is less than 0.42 dB, the resistance is below 5 Ω (Z < 5 Ω). An ideal capacitor would result in an attenuation of 0 dB.

Type:DC-Block 500
Frequency range:250 kHz – 500 MHz
Useable frequency range at -1 dB:50 kHz – 1 GHz
Port 1:50 Ω BNC jack
Port 2:50 Ω BNC plug
Max. voltage:500 V DC
max. RF-Power:5 W max.
Insertion loss
Input – Output
250 kHz – 500 MHz:
<0.42 dB
Dimensions W x H x D:90mm x 28mm x 35mm
Weight:ca. 100 g
According to standard:DIN ISO 11452-7

Datasheet DC-Block 500