FESP 5133-1330 – Circular Coil

Schwarzbeck FESP 5133-1330 Circular Radiating Loop
Schwarzbeck FESP 5133-1330 Circular Radiating Loop

The handheld coil FESP 5133-1330 was designed for immunity testing against magnetic fields up to approx. 20 kHz. The coil has 225 turns of Cu-wire on 15 layers. There are spacers to provide a constant distance of 50 mm between coil center and EuT-surface. The generated magnetic fieldstrength is direct proportional to the coil current. Typical applications are Immunity Tests according to the Measuring Method SF 01 G acc. to VG 95377 standard.

Medium coil diameter:126 mm
Medium Coil Distance to Surface:50 mm
Number of turns:225
Wire Diameter:2 mm
Mechanical Dimensions:190 x 190 x 290 mm
Terminals:4 mm Laboratory plug
Usable Frequency Range:0 – 20 (50) kHz
Inductance:6.5 mH
Resistance:0.5 Ω
Capacitance:240-300 pF
Resonance Frequency:120 kHz
Impedance:|Z|=42 Ω @ 1 kHz
Max. Current:20 A
Max. Fieldstrength at 50 mm spacing:17 kA/m
Max. magnetic fieldstrength at 1 A coil current in 50 mm distance:858.2 A/m
Required Current for 1000 A/m in 50 mm distance:1.165 A
Magnetic Flux Density at 1 A coil current in 50 mm distance:1.078 mT180.66 dBpT
Conversion Current-Magnetic Fielstrength:58.67 dB/m
Weight:5.3 kg
Standard:Measuring Method SF 01 G acc. to VG95377

Datasheet FESP 5133-1330
Security Notes FESP 5133-1330