The generated magnetic fieldstrength is proportional to the coil current. A distance ring allows a precise spacing of 50 mm between coil and EUT surface. Other distance rings can be used to scale the current-fieldstrength ratio. The highest fieldstrength levels can be achieved using the shortest ring. There are further rings available on request in order to provide a certain scaling between coil current and magnetic fieldstrength (e.g. 1 Amp coil current = 100 Amp/m fieldstrength).
Application as magnetic field probe:
The FESP 5132 can also be used to measure existing magnetic fields. The open circuit output voltage is directly proportional to the magnetic fieldstrength (at fixed frequency) or directly proportional to frequency (at constant magnetic fieldstrength).
Number of turns: | 20 |
Wire diameter: | 2 mm Cu (AWG 12) |
Maximum Coil Current: | 20 A (5 min.) |
Nominal Coil Current: | 10 A continuous |
Max. Magn. Field Strength: | 1500 A/m (5 min.) |
Nominal Magn. Field Strength: | 750 A/m continuous |
Magn. Field Strength, 1 A Coil Current:(with 50 mm distance ring) | 75.56 A/m (95 µT) |
Current required for 1 A/m:(with 50 mm distance ring) | 13.23 mA |
Medium coil diameter: | 120 mm |
Spacing coil center to measurement plane: | 50 mm |
Mechanical Dimensions: | 0.16 m x 0.25 (0.3) m |
Terminals: | 4 mm |
Usable Frequency Range: | 0 – 150 kHz |
Resistance: | 0.1 Ω |
Weight: | 1.2 kg |
Standard: | ISO 11452-8, MIL 461E/F/G RS101, EN 55103 5.18.3.2 |